MAHR inductive measuring probe P 2004 T, mechanical, +/- 2mm |PROMOTION
Brand: MAHR
Art.-no. MI39655605
Manufacturer article no. 5323011
EAN 4059192055059
Individual prices for customers after login
Application
Suitable for a wide variety of standard applications in multi-point measuring technology. Can also be used as an alternative to dial gauges or precision pointers in measuring instruments such as internal precision measuring instruments, precision pointer snap gauges, etc.
Execution
- Precise, ball-guided measuring pin
- Clamping shaft diameter 8 mm can be clamped along entire length
- Cemented carbide gauge slide M2.5 with 3 mm ball diameter
- 5-pin plug in line with IEC 130-9, cable length 2.5 m
Advantage
- Includes cable rerouting tool for easy conversion to radial cable outlet
- Excellent electromagnetic shielding
Delivery
Inductive measuring probe Millimar P2004, operating instructions, cap for side cable output, key for setting the forward stroke
Model | P 2004 T |
Compatibility length measurement scanner | TESA |
Max. spindle travel | 4.4 mm |
Measuring range of electronic length measuring technology | +/- 2 mm |
Spindle lift | Manual |
Measuring pin adjustment | Spring force |
Cable output direction | Axial, Radial |
Material of bellow | FPM natural fluororubber |
Measuring span of electronic length measuring technology | 4 mm |
Measurement force | 0.75 N |
Repetition standard deviation acc. to DIN 1319 | 0.1 µm |
Front end position of spindle | -2.2 mm to 0 mm |
LInearity deviation in the range +/- 0,5 mm | 0.4 µm |
LInearity deviation in the range +/- 1,0 mm | 1.5 µm |
LInearity deviation in the range +/- 2,0 mm | 3 µm |
Sensitivity deviation | 0.3 % |
Rear end position of the measuring bolt | +2.2 mm bis 4.4 mm |
Cable length | 2.5 m |
Mass force increase | 0.2 N/mm |
Measuring force tolerance | +/- 0.15 N |
Reversal error | 0.5 µm |
Select RAL-colour code
!! NOTE: On-screen visualisation of the colour differs from real colour shade!!